- 202 -
Tare the balance
Format:
T CR LF
Accessible responses:
T_A CR LF
T_D CR LF
- command understood and in progress
- command carried out
T_A CR LF
T_v CR LF
- command understood and in progress
- command understood but tarring range exceeded
T_A CR LF
T_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
T_I CR LF
- command understood but not accessible at this moment
Give tare value
Format:
OT CR LF
Accessible responses:
OT_TARA CR LF
– command carried out
Response format:
1 2
3
4-12
13
14 15 16
17
18 19
O T
space
tare
space
unit
space
CR LF
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
Caution:
Tare value is always given in adjustment unit.
Set tare
Format:
UT_TARE CR LF
, where
TARE
– tare value
Accessible responses:
UT_OK CR LF - command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognized (tare format incorrect)
Caution:
Use dot in tare format as decimal point.
Send stable measurement result in basic measuring unit
Format:
S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
S_I CR LF
- command understood but not accessible at this
moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit