1
Balance Analytical Scale
User Manual
Страница 1: ...1 Balance Analytical Scale User Manual ...
Страница 2: ...6 2 1 01Calibration weight 6 2 1 02 Calibration Step 6 2 2 Function setting 6 2 2 01Unit conversion 6 2 2 02 Speed adjustment 6 2 2 03 Stability adjustment 7 2 2 04 Counting Function 7 2 2 05 Factory setting 7 2 2 06 Percentage function 8 2 2 07 Peak hold 8 2 2 08 Baud rate optional 8 III DATA OUTPUT 9 IV FAULT JUDGMENT 10 V MAINTENANCE 11 PACKING LIST 11 ...
Страница 3: ...g The speed can be adjusted Easy to operate Full range tare clear accumulate overload underload display fault alarm within the full range Counting function Optional configuration and function expansion Model BAS 1004 BAS 1204 BAS 1204N BAS 2004 BAS 2204 BAS 2204N Max CAP 100g 120g 200g 220g Accuracy Readability 0 0001g 0 1mg Repeat ability 0 0002g Liner 0 0003g Stable time 3s Operation temp 5 35 P...
Страница 4: ...asy to cause temperature changes such as avoiding direct sunlight and away from heaters 6 Please avoid connecting with large machines or devices with interference and avoid the interference of other devices 7 The dry environment is prone to static electricity so corresponding measures can be taken to avoid the influence of static electricity on weighing 8 Keep the balance away from objects and equ...
Страница 5: ...5 1 4 Buttons and display The display and keyboard are like icons POWER SET TARE 0 T CONFRIM CAL UNIT Stability indicator Range progress bar Weighing Speed Stability UNIT Animal weighing ...
Страница 6: ... 1004 100g BAS 1204 100g BAS 2004 200g BAS 2204 200g 2 1 02 CAL STEP For BAS series balances please warm up for 1 hour and wait until the balance is stable before calibrating Take BAS 2004 as an example weighing no load 1 Press 0 to tare first the balance will display 0 0000g 2 Then press the CAL button the display shows CAL 200 flashing 3 At this time put a 200g calibration weight on the weighing...
Страница 7: ...nd you can use this for count the quantity And please keep the products have the same weight and the minimum weight must be 0 5mg The details operation as blow 1 Press SET to enter menu then press SET unit it shows COU 2 Press O T to enter it shows COU 5pcs 3 Press UNIT to choose sample quantity put matched quantity press O T it shows 4 When COU disappearing showing digital then you can operate 5 ...
Страница 8: ...07 PEAK HOLD The steps are as below 1 Press SET enter menu then press SET until it shows PEAK 2 Press O T enter peak hold function it shows ON OFF press UNIT to choose press O T to confirm ON OFF 2 2 08 BAUD RATE CHOOSING The steps are as below 1 Press SET to enter menu then press SET until it shows BAUD 2 Press O T to enter baud rate choosing it shows 1200 press UNIT to choose what you need press...
Страница 9: ...ud rate 1200 2 Data bits 8 3 Parity No 4 Stop bit 1 5 Data flow control hardware Click OK display the current weighing value 3 3 Balance data output mode setting The output mode of the balance is continuous 3 4 The output sequence of one frame data of the balance is as follows 1 Frame header E Fixed 2 Space 0X20 Fixed 3 stable state unstable Space stable 4 Symbol Space 5 13 Weighing data Decimal p...
Страница 10: ...r evaporating Plastic particles or powders are electrostatically charged 6 Static electricity generated by dry weather Improve the working environment and avoid the interference of vibration and airflow Close the windproof door Take out the foreign body turn the weighing pan to prevent scratching External access 9V DC power adapter Operate with static eliminator or wristband with static electricit...
Страница 11: ...and fine particles please use it with the container to avoid dust and particles falling into the load cell under the weighing pan 6 Please wear gloves when calibrating with weights and do not take the weights directly with your hands 7 Keep the balance clean and dry Matters needing attention when cleaning Before cleaning unplug the power supply Do not use corrosive cleaning agents such as solvents...