- 75 -
18.
DOSING
Working mode
<
Dosing>
enables carrying out the sampling process, until
obtaining a pre-defined target mass.
Working mode activating procedure
•
while in the main window, press
soft key located in the upper bar of the
display, which opens a submenu <Working modes> with selection of
available working modes,
•
select <
Dosing > mode, the software automatically returns to the main
window and displays
pictogram in the upper bar.
The grey coloured workspace contains the following data:
o
Tare
o
Gross
o
Target value
o
Tolerance – percent [%] value related to the target value
o
Product
On selecting the dosing mode, the display contains the following quick
access keys in the bottom bar:
1.
Setup – access to balance’s menu
2.
Print header – print of data declared in the header
3.
Print footer – print of data declared in the footer
4.
Database of products – selection of products from corresponding
database
5.
Set tare – field for setting numeric value of tare
6.
Target value – for declaring the Target value of dosing process