![RADWAG 3Y Series Скачать руководство пользователя страница 197](http://html1.mh-extra.com/html/radwag/3y-series/3y-series_user-manual_3937928197.webp)
- 197 -
Example of a report:
.....................................................
--------- Autotest GLP: Report ---------
Balance type PS 3Y
Balance ID 400010
User Admin
Software rev. L0.0.21 S
Date 2012.01.16
Time 09:17:16
----------------------------------------
Number of measurements 10
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
Filter Average
Value release Fast & Reliable
----------------------------------------
Deviation for Max. -0.118 g
Repeatability 0.0088 g
Signature
........................................
PROCEDURE:
On pressing a field with test name, the balance displays a message box.
On this level, the user can carry out the following:
Start the following AUTOTESTU GLP process
Preview results of completed autotests, and export all saved test results as
file with extension *.csv
Delete all saved test results
User can abort the process at optional moment by pressing <X> key in the
testing window.
Results of carried out autotests are displayed as a table (each line features date
and time of autotest procedure and its result).
To preview detailed data of an autotest, press a specified line of the table with
test results.
In order to print a result of a single autotest, enter its details (see above) and
generate a printout by pressing <print> key in the upper bar of the balance’s
display.
Autotest results can be exported by pressing <export> field on the level of all
saved reports in a table. Data is sent to a data storage device connected to
balance’s USB port as a file with extension *.csv.
Содержание 3Y Series
Страница 2: ...2 APRIL 2014...
Страница 21: ...21 3 BALANCE DESIGN 3 1 Dimensions Ulctramicrobalance UYA 3Y series Balance XA 3Y A series...
Страница 22: ...22 Balance XA 3Y series Balance XA 3Y F series...
Страница 23: ...23 Balance AS 3Y series Balance PS 3Y series...
Страница 24: ...24 Balance APP 3Y series...