Autozero function: Dosing
The function sets autozero operation mode to default one set for
dosing mode.
Accessible values:
NO
- autozero operation is automatically disabled on entering
dosing mode
YES
- autozero operation is automatically enabled on entering
dosing mode
Last digit
The function determines visibility of the last decimal place
indicated on measurement result. The functions provides three
settings:
Ambient conditions
this parameter refers to ambient conditions of the balance. By default the
parameter is set to value
Stable
but it is advisable to set the parameter
to value
Unstable
for unstable ambient conditions.
Stable
setting results
in faster operation, i.e. weighing takes much less time than in case of
setting to value unstable.
−
Unstable
−
Stable
14.2.4.
Measuring units
Balance user can declare in a profile: start unit, supplementary unit and two
custom measuring units.
A custom measuring unit features:
o
A multiplier
o
A name (3 characters)
If a custom unit is designed, then its name is added to the list of accessible
measuring units.
This menu additionally enables inserting the value of gravitational acceleration
force for balance’s place of use. It is obligatory should a balance be used to
determine mass in [N].