27.10.
Destructive empty-full and full-empty control mode
In the
“destructive”
control mode independently on the size of product lot over
100 pieces, the sample set by the software for testing is always 20 pieces. The
other control assessing criteria are set in accordance with the regulation.
After using the list to select a product with set parameters for destructive control
with determined measuring
“match”
(manufacturing unit) and starting a
destructive control, the software displays commands aiding the control process
(as in case of the controls described above).
Depending on enabled control mode, the sequence of products to be weighed is
either:
“empty-full”
or
“full-empty”
.
Caution:
Remember to maintain the weighing sequence of products together with the
packages and empty packages. It is necessary to let the software correctly
calculate mass of a load placed in its corresponding packaging.
On control completion, the software generates a process summary and the
completed control is automatically saved in the balance’s database.
Caution:
A template and an example of a report on product control is provided further
down this user manual.
27.11.
Control according to internal criteria
Select appropriate product with correctly entered data, when the data is entered
in regard to control performed according to internal criteria (see chapters
above).
Press button
to start control after weighing mode parameters have been
entered into balance memory (accordingly to the instruction). The button is
located in a bottom part of a window. In course of the control the software
continuously analyses measurements results and displays them in a respective
fields: