- 209 -
Format of mass printout:
1
2
3
4 -12
13
14 15 16 17 18
stability
marker
space character mass space
unit
CR LF
Stability marker
[space] if stable
[?] if unstable
[!] air buoyancy compensation mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to – occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification
Example:
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF
– a printout generated from a balance
on pressing
key, according to settings of <Weighing printout
template>:
N (number of
measurements)
NO
Universal variable 1 … 5
NO
Date
NO
Net
NO
Time
NO
Tare
NO
Level status
NO
Gross
NO
Client
NO
Current result
NO
Warehouse
NO
Supplementary unit
NO
Product
NO
Mass
YES
Packaging
NO
Non-standard printout
NO