- 147 -
- Nominal value of a controlled product
- Average mass of a controlled product
- Value of disqualifying average
- Negative errors
T
characteristics per sample:
-9g
– value of negative error
T
,
T Max
– tolerable number of negative errors
T
,
n
- actual number of negative errors
T
- Negative errors
2T
characteristics per sample:
-18g
- value of negative error
2T
,
2T Max
- tolerable number of negative errors
2T
,
n
- actual number of negative errors
2T
- Net mass of a controlled product
- Packaging tare
- Control status: positive, negative
Load full
- Command for process in progress with specified
number of all packagings to be weighed
- Work pane options: numerical data / graph
- Information about control or performed control
measurements indications
- Abortion of a control process
•
Control status
Control status is presented by the following pictograms:
- positive,
- negative (control of sample 2 permitted)
- negative
For status
respective field of a work pane turns to yellow:
- yellow warns that occurrence of another
T
error
will end with negative result of control.
For status
respective field of a work pane turns to red:
- average mass of controlled product is lower than
disqualifying average value