Titra
Lab
®
TitraLab 870 Reference Manual
Page 74
Calibration
stack
For an electrode calibration method,
.
For a reagent calibration method,
see "Reagent calibration stack",
.
Catalogue list
List of Radiometer Analytical names of electrodes and commonly
used reagents. This list cannot be modified.
Cell constant
(parameter)
Enter the cell constant value. The cell constant is a specification of the
conductivity cell and is normally provided by the cell manufacturer.
If you do not know the cell constant value or if you want to
check its value, select
Calibration request =
Yes
, edit and run a calibration method.
It is recommended to periodically check the constant value by
performing a cell calibration.
Refer to "Electrode calibration (conductivity)", page 108
.
Refer to "Cell constant (definition)", page 75
Access:
Edit Electrode menu (for a Conductivity type of electrode with
Calibration request = No
)
Range available:
0.050 to 15.000 cm
-1
(by steps of 0.001 cm
-1
)
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...