Dynamic rinses
Dynamic rinses are performed by a SAC850 or SAC950 Sample
Changers if the Dynamic rinsing module is installed on the sample
changer.
A Dynamic rinse performed in a Park or Rinse beaker consists of the
following operations:
1.
The electrode are positioned above the Park or Rinse beaker.
2.
The electrodes are dipped into the beaker. The beaker is emptied
to a waste in the same time. At the end, the electrodes are
located in the emptied beaker at their downmost position.
3.
The electrodes are washed with rinse solution (usually demineral-
ised water) then start to move up under rinsing. At the end, the
electrodes are located above a beaker filled with rinse solution
and some remaining impurities.
4.
Steps
2
and
3
can be repeated up to 8 times as up to 9 dynamic
rinses can be programmed. Steps
2
and
3
are performed under
stirring.
A Dynamic rinse performed in a Sample beaker consists of the follow-
ing operations:
1.
The electrodes are dipped into the Sample beaker. The beaker is
emptied to a waste in the same time. At the end, the electrodes
are located in the emptied beaker at their downmost position.
2.
The electrodes move up and are washed in the same time with
rinse solution (usually demineralised water). At the end, the elec-
trodes are located above a beaker filled with rinse solution and
some remaining impurities.
3.
The sample beaker is emptied to a waste.
4.
A last rinsing of the electrodes and delivery tips is carried out. At
the end, the electrodes are located above a nearly emptied
beaker containing a little solvent that was used to flush the elec-
trodes and delivery tips.
5.
Steps
1
to
3
can be repeated up to 8 times as up to 9 dynamic
rinses can be programmed. Steps
1
to
3
are performed under
stirring.
E0 standard
potential
Refer to "Direct ISE measurement method - definition", page 95
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...