ISE Standard
addition method
-
programmation
Proceed as follows to edit an ISE standard addition method:
1.
From the Main window, press
4
then
2
Edit method.
2.
For
Mode
, select
Measurement
.
3.
For
Measurement
, select
ISE Std Addition
.
4.
Use the LEFT/RIGHT arrow keys to move to the last Edit method
display.
5.
Press
1
Method parameters.
6.
Select the ISE electrode, the standard addition ID and the number
of additions to be performed (from 1 to 9).
7.
At the line Std add volume:
•
select
Programmed
then enter the volume of one addition
(in ml),
•
or select
Automatic
then enter the potential jump due to all n
additions (n = 1 to 9).
The TIM870 calculates the volume to be poured for each addition.
To avoid modifying the ionic strength of the solution by dilution
effect, it is recommended to add small amounts of standard.
The addition cumulate volumes should not exceed 10% of the
sample test quantity.
In an ISE Standard addition method with automatic
determination of the volumes, if the cumulate volumes
calculated is higher than 50 % of the sample test quantity, the
analysis stops and the instrument calculates the results with
the data available. A
Vstd>max
note is added to the result
value to show that the cumulate volumes of the standard
additions is higher than 50 % of the sample test quantity.
8.
Define the other parameters of this measurement method.
What is an ISE Standard addition measurement method?
see "ISE Standard addition method - definition", page 142
How to run an ISE standard addition method?
Refer to "Running a method", page 216
.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...