TitraLab 870 Reference Manual
Page 82
Continuous IP
method
For this type of method, the titration is carried out using continuous
addition of titrant.
The titrant addition speed is regulated between two user selectable
limits (Min. speed and Max. speed). The addition speed decreases
when the slope of the E or pH = f(Volume) titration curve increases
and vice versa. This prevents overshooting the equivalent point.
Figure 15: Continuous IP - Titrant addition speed
Using a Continuous IP method, higher the slope of the titration curve
is, higher the number of measurements saved will be. A great majority
of points are saved close to the inflection point. This allows an accu-
rate determination of the equivalent point.
Figure 16: Continuous IP - Number of points saved
The Continuous IP method can be stopped after a given number (1 to
8) of equivalent points has been detected.
Speed
Volume
V max
V min
E or pH
E or pH
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...