Programming tips
•
Do not forget to declare electrode(s) while setting up reagent parameters, if calibrated.
•
Do not forget to declare electrode(s) and reagents(s) when programming your method
parameters
•
If reagents are calibrated, the same electrode must be used in both sample analysis
method and reagent calibration.
•
If a Sample Changer is used, do not forget to declare one in the Configuration menu.
If no sun icons appear after the method has been selected, check the following points:
1.
Install electrode(s) for selected method,
see "Check electrodes", page 75.
2.
Install reagent(s) for selected method,
see "Install a reagent in a burette stand", page 128.
3.
If required, calibrate electrode,
see "Calibrate pH electrodes", page 69.
4.
Calibrate/enter reagent titre,
Refer to "Calibrate reagents", page 69
Refer to "Enter titre", page 111
If Sunny icons appear
These are required to run the selected method.
If the Cloudy icon appears
:
The electrode/reagent calibration should be performed within 24 hours. The expi-
ry date of one of the reagents in the system will expire in less than one week.
This is a simple warning, it will not stop you from running the analysis.
If the Stormy icon appears
:
Titrant and/or electrode required in the selected method is/are not installed or
have not been calibrated.
If the Question mark icon appears
:
It is a programming error, reagent and/or electrode is/are not defined in the
selected method. Revise the method programming.
If the Calculator icon appears
:
Reprocessing mode (Working mode = Reprocessing) is set on your instrument.
Change the working mode if you want to run a method or a sequence.
Refer to "Working mode", page 235
.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...