TitraLab 870 Reference Manual
Page 151
Measurement
Measurement type for the method.
Enter in:
Edit method menu.
Edit reagent menu if
Titre = Calibrate
.
Range available:
pH measurement (pH), zero-current potential measurement (mV),
imposed current potential measurement (mV(i>0)).
For Measurement methods (Mode = Measurement :
), 3 other options are available : Direct ISE, ISE Std Add and
Conductivity.
If you select
mV(i>0)
, connect the double platinum electrode to
the
Pt-Pt
socket on the rear panel. One of the electrode's poles
is connected to the ground, so it is necessary to select
TIM
ext. cell Gnd = No
in the Setup window.
If you select
Conductivity
, connect the conductivity cell to the
EC
socket on the rear panel. One of the
EC
socket pin is
connected to the ground, so it is necessary to select
TIM
ext. cell Gnd = No
in the Setup window.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...