Titra
Lab
®
TitraLab 870 Reference Manual
Page 55
Auxiliary output
The auxiliary outputs are used to control external equipment such as
valves or pumps during titrations. This signal is sent between the red
and black banana sockets
5V OUT
or
12V OUT
of the Titration
Manager.
Auxiliary output (5 V, 12 V, No)
Activate to 5 V or 12 V or disable the auxiliary signal.
Specifications of the auxiliary ouput signal:
see "TTL 5 V OUT/TTL 12 V OUT (sockets)", page 250
Aux. action
Titration start
The auxiliary signal is initiated at titration start (before predose if rele-
vant). Duration is set by
Aux. on for
.
Titration end
The auxiliary signal is initiated at titration end. Duration is set by
Aux. on for
.
Whole titration
The auxiliary signal is initiated at titration start (before predose if rele-
vant), and lasts the whole titration.
Aux. on for
Time during which the auxiliary signal is set to 5 V or 12 V when the
Titration start
or
Titration end
is selected for
Aux. action
.
Enter in:
Method parameters menu
Reagent parameters menu
Range available:
Aux. action: Titration start, Titration end or Whole titration
Aux. on for: 0 to 99:59 min:s
Measurement and Addition methods:
The
Aux. action
parameter is not available in a
Measurement or an Addition method. An auxiliary output can
be activated:
•at the measurement start or before the first reagent addition
(duration set by
Aux. on for
)
•
or during the whole measurement including measurement
stabilisation delay or during all the reagent addition(s). In this
case, select a 5 V or 12 V auxiliary output and set
Aux. on for = 0
.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...