Titra
Lab
®
TitraLab 870 Reference Manual
Page 233
Solution menu
This menu is available for pH or ISE electrodes that are edited with
the
Calibrate request = Yes
option.
To access:
1.
From the Electrode window, press
4
.
2.
Select the electrode to be edited.
3.
Press
2
Edit electrode and check that the
Calibration
request = Yes
option has been selected.
4.
Use the
LEFT/RIGHT
arrow keys to move to the last
Edit electrode display.
5.
Press
2
Calibration solutions.
For a pH electrode calibration
For an ISE electrode calibration
Select the pH standard solutions to be used for
the calibration. The pH values are given at
25°C. The following values are available: IUPAC
pH standards (pH 1.679, 4.005, 6.865, 7.000,
7.413, 9.180, 10.012 or 12.454) or 4-7-10
Series (pH 4, 7 or 10).
Select the standard used from the User or the
Catalogue list.
Select the standard concentration unit.
Available units: eq/l, meq/l, mol/l, mmol/l, g/l,
mg/l, mg/ml, µg/ml, % or ppm.
Enter the concentration of each standard with
the unit selected above.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...