Titra
Lab
®
TitraLab 870 Reference Manual
Page 155
Min. ordinate
Max. ordinate
Set the lower and upper limits of a search interval for a given inflection
point (IP). Min. and Max. ordinates are entered for each IP to be deter-
mined.
Figure 28: Minimum and maximum ordinates
Only one inflection point (IP) is determined for each Min./Max. ordinate
interval.
When several IP are found in the same interval:
•
The first IP is the result,
•
The other IP are rejected and a note “
IP>1
” is displayed.
If you know in advance the pH or mV of the inflection point, define a
search interval corresponding to the expected value.
Enter in:
Parameters menu of a Monotonic IP, Dynamic IP or Continuous IP
method.
Range:
Min. ordinate: -9.000 pH (or 2000 mV) to Max. ordinate
Max. ordinate: MIn. ordinate to +23.000 pH (or -2000 mV).
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...