TitraLab 870 Reference Manual
Page 112
Electrode
calibration not
required
Message appears at the start of a sequence, if a method sequence
has been programmed with an electrode calibration. The electrode
used has been programmed without calibration
Calibration
request = No
.
Go to Sequence/Sample stack,
Edit sequence
menu and
remove the electrode/reagent calibration method.
Electrode
calibration
parameters
This menu contains the general parameters concerning the
electrode calibration method (measurement stabilisation criteria in
particular).
To access:
1.
From the Electrode window, press
4
.
2.
Select the electrode to be edited.
3.
Press
2
Edit electrode and check that the
Calibration
request = Yes
option has been selected.
4.
Edit the electrode calibration general parameters.
5.
Use the
LEFT/RIGHT
arrow keys to move to the last Edit
electrode display.
6.
Press
1
Calibration parameters.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...