Electrode icons
Select to access Electrode window.
Indicates the state of the electrode system.
Sunny icon
:
The calibration has been performed on all the electrodes
present in the system and/or all the electrodes have been
installed.
Cloudy icon
:
The electrode calibration of one of the electrodes present
in the system should be performed within 24 hours.
Note: when the Periodicity is set to 1 day, this icon will
appear to indicate that a calibration must be performed
within 12 hours.
Stormy icon
:
The calibration date has elapsed for one of the electrodes
present in the system.
If acceptance limits have been set for the calibration: at
least one calibration result lies outside the programmed
acceptance limits.
At least one of the electrodes present in the system has
not been installed.
Question mark
:
The electrode system has not been programmed
correctly. Enter Supervisor mode and Check the
electrode parameters in the Method parameters menu. If
a temperature sensor has been defined in the Electrode
menu, use the same sensor in method.
For a reagent calibration, make sure that the electrode(s)
used for the calibration are the same as those used in the
method.
Calculator icon:
Reprocessing mode (Working mode = Reprocessing) is
set on the instrument. This icon is displayed irrespective
of the status of the electrode system used.
Refer to "Working mode", page 254
Press
1
in the Main window, the instrument will indicate the
possible errors and prompt you to correct them.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...