TitraLab 870 Reference Manual
Page 254
Working mode
Select the way in which you want to work.
Method
: to run a single method.
Sequence
: to create or run a sequence of methods. Beakers are
manually changed between two method runs.
SAC Method
: to run a single method to be performed using a sample
changer.
SAC Sequence
: to create a sequence of methods to be performed
using a sample changer.
Reprocessing
: to reprocess in manual mode the last titration curve
obtained.
Note: the working mode selected will have no effect on the type of
method you wish to create.
Define the sample changer in the Configuration menu before
selecting SAC Method or SAC Sequence.
Enter in:
Main window
Wrong buffer
While running an electrode calibration, the buffer analysed has not
been identifed by the instrument.
End the calibration cycle. Check the buffer then start a new calibration
cycle.
Wrong type
(SAC error)
Sample changer error: the sample changer connected does not corre-
spond to the one programmed in the Setup > Configuration menu.
Example:
A SAC850 is connected to the Titration Manager,
SAC
socket and you
have declared a SAC950 in the Setup > Configuration menu.
Review the Setup > Configuration parameters and/or connect the
correct model of Sample Changer to the SAC socket of the Titration
Manager.
Refer to the Sample Changer User’s Guide, part no. D21T085.
Zero pH
This is the pH value (pH
0
) at which the potential is zero. The zero pH
is calculated after a one-point calibration.
.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...