TitraLab 870 Reference Manual
Page 141
IP reject
The IP reject sets lower level of the slope of the titration curve for
acceptance of an inflection point. Use the IP reject value to eliminate
weak inflection points.
To set the appropriate IP reject value, display the derivative curve of
the titration and find the IP reject value as shown below:
Figure 25: IP reject
Enter in:
Parameters menu of a Monotonic IP or Dynamic IP method.
Range available:
1 to 999 (unit = mV/ml or pH/ml)
IP>1 (result
indicator)
Refer to "Result indicators", page 206
ISE calibration
results
parameters
Refer to "Results menu", page 210
ISE calibration
solutions
Refer to "Solution menu", page 233
IP reject 2
d Potential (or d pH)/dVolume
IP reject 1
Volume
IP reject 3
IP reject 4
IP1
IP2
IP3
IP reject 1: 3 Inflection points found (IP1, IP2 and IP3)
IP reject 2: 2 Inflection points found (IP2 and IP3)
IP reject 3: 1 Inflection point found (IP3)
IP reject 4: no inflection point found
Direct curve
Derivative curve
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...