Results by
difference/
cumulate
Results by difference (ml)
Volumes between two End points or Inflection points are used for
result calculation.
Results by cumulate (ml)
Total volume delivered up to the End point or Inflection point is used
for result calculation.
This setting has no influence on the result calculation for the
first equivalence point detected.
Enter in:
Edit method > Results (End point, IP monotonic, IP dynamic or IP
continuous methods)
Results factor
(Yes/No)
This multiplying factor is applied to all Rn (n = 1 to 8) results of the
method. Equation results En (n = 1 or 2) are not processed.
Access:
Edit method > Results
Proceed as follows to multiply all your method results by a factor:
•
In the method Results menu, select
Results factor =
Yes
.
•
Run the method:
There is no results factor for a reagent calibration method.
Enter the results factor (from 0.001 to 10 000)
then press
1
to start the method.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...