Max. stab time
If the stability criterion has not been fullfiled during the time entered for
the Maximum stabilisation time an error message will appear. Check
your electrode before repeating the measurement.
Enter in:
Edit method > Parameters menu (Measurement method only)
Edit electrode > Calibration parameters menu
Range available:
0 to 59:59 min:s
Max. volume
This is the maximum volume of titrant to be added during the titration.
The titration will stop once this volume reaches the entered value.
A burette cannot be refilled during a Continuous IP method.
So, while editing an Continuous IP method, enter a
Max. volume
so that you have:
Max. volume - Predose until
< burette volume.
An error message is displayed (Max. vol - Predose > Bur. vol)
if this condition is not fulfilled and you cannot start the method.
Enter in:
Edit method > Parameters menu
Edit reagent > Parameters menu
Range available:
Predose until
(in ml) to 999.9 ml
In Continuous IP:
From
Predose until
(in ml) up to the burette volume.
Max. vol -
Predose > Bur.
vol
This error message indicates that a burette cannot be refilled during a
Continuous IP method. You cannot run the method.
Press
✓
and edit the method with a
Max. volume
so that you
have:
Max. volume - Predose until
< burette volume.
or install a burette of a sufficient volume.
Max. volume
reached
In a titration method with a predose to set pH or mV option, the
potential or pH has not reached before the Max. volume.
Resume the test or end the analysis.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...