TitraLab 870 Reference Manual
Page 113
Electrode
calibration
stack
The electrode calibration stack defines individual data for each buffer
solution beakers present in an electrode calibration sequence.
1.
In the Configuration menu, declare the model of sample changer
used (SAC80, SAC90, SAC850 or SAC950).
Refer to "Configuration menu", page 81
Depending on the model of sample changer used, enter the
sample changer configuration parameters.
Refer to "Sample changer", page 221
2.
In the Main window, select
SAC Sequence
or
SAC method
for the
working mode
. This sequence or method must use
the electrode you want to calibrate. Edit the sequence or the
method if relevant,
see "Programming sequence", page 181
3.
Enter the Electrode window and press
1
Calibrate electrodes.
4.
Select the electrode to calibrate.
5.
Press
2
Calibration sequence.
Label the beakers indicating the running number in the
sequence, for example: 1/15, 2/15 etc.... and the name of the
buffer solution.
Place the beakers in the numbered position on the sample
changer.
If rinses are programmed, position the corresponding rinse
beakers at the right places.
Refer to "Number of static rinses", page 166
.
Refer to "Dynamic rinses", page 102
.
You can print the calibration stack by pressing
from the
calibration menu.
6.
Press
Esc
then run the sequence by pressing
1
Run calibration.
<1/15> means the first beaker over 15
programmed in the sequence. Use the
LEFT/
RIGHT
arrows to review the other beakers in
the sequence.
Run 1/5 means that this beaker deals with the
first cycle over 5 programmed in the
sequence.
Enter the batch number of each buffer
solution.
Beakers are numbered in that order:
Cycle 1, Buffer 1 - Cycle 1, Buffer 2 ..... Cycle 1, Buffer n (n=1 to 5)
Cycle 2, Buffer 1 - Cycle 2, Buffer 2 .....Cycle 2, Buffer n
.......................................................
Cycle m (m=1 to 9), Buffer 1 - Cycle m, Buffer 2 .....Cycle m, Buffer n
n and m are entered in the Edit electrode menu. The buffer solutions are selected
in the Solutions menu.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...