Page 84 of 87
Service & Calibration
5.3.1 Battery
Replacement
Backup power for the instruments non-volatile memory is supplied by 3 standard AA
batteries. These should be
replaced annually with alkaline batteries,
otherwise damage
to the unit is possible as a result of battery leakage. QuadTech will assume no
responsibility for instrument damage resulting from the batteries not being changed as
recommended. To replace the batteries proceed as follows:
1. Instrument power needs to be on but not in the measurement mode. If the batteries
are removed with the power off there would be no power to the non-volatile memory
and important data such as stored test conditions would be lost.
2. Place a flat head screwdriver in the groove of the battery compartment end cap and
rotate counterclockwise (CCW) about 1/4 turn to remove.
3. Lift the front of the instrument up slightly so that the batteries slide out of the
compartment.
4. Install three new batteries,
positive side out.
CAUTION
Do not install the batteries backwards
5. Reinstall the end cap and rotate clockwise (CW) to secure.
IEEE-488 INTERFACE
I/O PORT
PARALLEL PORT
RS-232 INTERFACE
BATTERY
90 - 250 V
47 - 63 Hz
40 WATTS MAX
Battery Compartment
Figure 5-1: Battery Compartment