Operation
Page 43 of 87
2.6.2.8 Limit
Setup
= 10
= 10
Dwell Time
Measure Time
Voltage
I/O Utilities
= 1000
e
Stop on Pass = 0
to Average = 0
#
10
Discharge Time
Range > >
Limit
=
=
1 e + 006
Charge Time = 10
Mode Auto Manual
In the case where I (current) is selected rather
than R (resistance), the exponent would be
entered
as
a
negative
number.
Figure 2-15: Limit Value (Resistance or Current)
Allows entry of a single measurement limit in scientific units (up to 4 digits with decimal
and exponent) for resistance or current depending on the results parameter selected. The
exponent must be 3 through 14 for resistance or -3 through -13 for current. This limit
determines the status of the Pass/Fail indicator and Pass/Fail display, when enabled.
When the 1865 instrument is in the resistance mode the limit is a minimum value i.e. any
value above the limit will result in a Pass, whereas a value below the limit will result in a
Fail. The opposite is true when the unit is in the current mode, the limit is a maximum
value, any value below the limit will result in a Pass and a value above the limit will
result in a Fail.
If the display parameter on the I/O Menu is selected as resistance the value entered will
be an R limit, if the display is selected as current the value entered will be an I limit. The
1865's processor will check for consistency between the parameter display selected and
limit entered.
An example of the sequence for entering 1M
Ω
as a limit would be:
Press
[ENTER]
or
Press
[ENTER]
Press
[1]
Press
[1]
Press
[e]
select
key
Press
[0]
(zero), three times
Press
[6]
(exponent
multiplier) Press
[e]
select key
Press
[ENTER]
Press
[3]
Press
[ENTER]