Page 72 of 87
Operation
"PASSWORD NOT SAVED IN RAM"
Error trying to save password, repeat
"PASSWORD VERIFICATION FAILURE"
Wrong password entry for verification
"RANGE RESISTOR VALUE OUT OF RANGE" Malfunction of internal calibration resistors
"REMOTE COMMAND INVALID"
IEEE or RS232 command is incorrect (for
example: to set voltage use VOLTage)
"REMOTE COMMAND PARAMETER
IEEE or RS232 parameter is incorrect (for
INVALID"
example: voltage parameter can be
XXXX.XXX)
"REMOTE COMMAND PREFIX INVALID"
IEEE or RS232 command prefix is
incorrect (for example: to configure unit
use
CONFigure)
"SHUT THE POWER OFF"
Malfunction of high voltage source
"TIME / DATE ERROR, PLEASE SET"
Reset time and date
"UNABLE TO READ FROM FLOPPY"
Floppy option not present, defective or disk
not
inserted
"UNABLE TO READ THAT FILENAME"
Selected file unreadable, improper format
"UNABLE TO SAVE CALIBRATION DATA"
Repeat calibration, data not stored
"UNABLE TO WRITE TO FLOPPY"
Floppy option not present, defective or disk
not
inserted
"UNABLE TO WRITE TO REMOTE"
IEEE or RS232 not setup correctly or IEEE
option not present
"UNKNOWN UNIT"
Floppy drive is not recognized (option not
present or defective)
"VALID RANGE = low value - hi value"
Entry invalid, should be between specified
values
"VOLTAGE PROGRAM FAILURE"
Malfunction of instrument power supplies