Applications
Page 77 of 87
3.4 Resistance
Measurements
The recommended test voltage is typically 100V for fixed composition resistors, film
resistors or wire wound resistors above 100k
Ω
. Refer to EIA (Electrical Industry
Association) Standards. These resistors (values greater than 1000 ohms) can easily be
measured on the 1865 where the instrument basic accuracy of 0.5% is adequate. For
single component resistors a two terminal connection, +UNKNOWN and -UNKNOWN
is recommended.
3.5
Measurement of Voltage Coefficient
The 1865 instrument may be used to measure voltage coefficient which is defined as :
R1
-
R2
----------------- x 100%
R2 (V1 - V2)
where V1 > V2
R1 is the resistance at V1, the higher voltage
R2 is the resistance at V2
For example, if V1 = 500 V and V2 = 100 V,
R500V - R100V
Voltage Coefficient = --------------------- x 100%
R100V (400)
1
∆
R
= --- ------- %
4
R100V
This voltage coefficient is usually negative.
Содержание 1865