Page 19 of 87
Condensed Operating Instructions
Setting Pass/Fail Limit
1. To enter a single measurement limit for resistance or current (depending on results
display selected):
Press [MENU] key to select menu display.
Press Right or Left Arrow to select Setup menu.
Press Up or Down Arrow key to select Limit in sub menu.
Press [ENTER] key to activate the limit entry field.
Enter the numerical value
(up to 4 digits plus decimal point) of the limit
desired, then enter the exponential value
after first pressing the
e
select key
(use minus exponential value for current).
Press [ENTER] key to finalize the limit entry.
Changing Display Type and Units
1. To change displayed measurement results between Resistance, Current or Pass/Fail:
Press
[MENU]
key to select menu display.
Press Right or Left Arrow key to select I/O Menu.
Press Up or Down Arrow key to select Display.
Press [ENTER] key to activate the selection field.
Press Up or Down Arrow key to select desired measurement results,
Resistance, Current or Pass/Fail if a limit is to be used.
Press [ENTER] key to finalize the entry.
2. To change the measurement results format between Scientific or Engineering units:
Press [MENU] key to select menu display.
Press Right or Left Arrow key to select I/O Menu.
Press Up or Down Arrow key to select Result Format.
Press
[ENTER] key to activate the selection field.
Press Right or Left Arrow to select desired results format, Sci for Scientific or
Eng for Engineering.
Press
[ENTER]
key to finalize the entry.