Page 34 of 87
Operation
2.5.3 Automatic Measurement Mode
Measurements in the automatic mode can be initiated after connecting the DUT by
simply pressing [START]. Test conditions are determined by either the power up default
conditions or recalling a previously stored setup from instrument memory. In either case
a test cycle is sequenced automatically (Figure 2-5) once [START] is pressed and results
displayed (Figure 2-6). When a test limit is entered a PASS/FAIL can be displayed in
place of the actual measured value (Figure 2-7).
Charge
Vo ltage
applied
to UUT
AUTOMATIC M EAS UREM ENT
MODE
STA RT
Test Cycle Time (P rogra mmab le )
(
H igh Voltage
Indicator On)
(H igh Voltage
Indicator Off)
Dis charge t
(0 - 300s )
M easure t
(0 - 300s )
or (0 - 999)*
Dwell t
(0 - 300s )
Charge t
(0 - 300s )
* on lat er unit s
Figure 2-5: Test Cycle Sequence (Automatic)