Page 54 of 87
Operation
When there are more setups than can fit on the display the page down key is active. If
there is less than a whole page below, the display wraps around to the previous display.
Continuing to page down will eventually return to the first display of setups.
NOTE
Setups saved using version 1.6 (or earlier) operating software can not be recalled with
software 2.0 or later. The software version is shown on the display at power up.
2.6.4.3 Zero
The zeroing process automatically measures stray parameters and retains the data, which
is used to correct measurements so that results represent parameters of the DUT alone
without test lead or fixture capacitance. Zeroing is recommended at the start of each
work day or more often if leads, fixture or test configuration to the DUT is changed.
Zeroing should also be performed anytime the test voltage is changed, which also
includes recalling a set of test conditions from memory with a different test voltage. It is
important to note, that anytime the instrument is zeroed it is done at the test voltage
currently specified. Once [Zero] is selected in the menu and [Enter] key pressed the
operator is prompted by instructions on the display, the first is shown in Figure 2-26.
REMOVE ALL COMPONENTS
FROM THE TEST FIXTURE
HIT <START> WHEN READY
HIT <CNCL> TO EXIT
ZERO AT 500 VOLTS
Figure 2-26: Zeroing Procedure
Once the zeroing is initiated by pressing [START] the next message displayed is:
ZERO
CAL
IN
PROGRESS
Once complete, typically a minute or slightly longer, the next message displayed is:
COMPLETE
HIT <MENU> KEY TO CONTINUE