Operation
Page 37 of 87
2.6.2 Setup
Menus
Setup
I/O Utilities
Voltage
Charge Time
Mode Auto Manual
Dwell Time
Measure Time
Discharge Time
= 0
= 0
= 0
Range > >
Limit
= 1000
= 0
= 100 e+ 9
Stop on Pass = 0
to Average = 0
#
Figure 2-9: Setup Menu
The first of the three main menus is Setup, shown in Figure 2-9. Each function controls a
measurement condition and is described in detail below.
2.6.2.1 Voltage
e
Charge Time = 10
Setup
= 10
= 10
= 10
Dwell Time
Measure Time
Discharge Time
Mode Auto Manual
Voltage
I/O Utilities
Range > >
Limit
1000
= 100 e+ 9
=
Stop on Pass = 0
to Average = 0
#
Figure 2-10: Test Voltage Entry
Accepts entry (up to 6 digits and decimal) of a test voltage between 1 and
1000 volts. This is the voltage applied to the unit under test during the test cycle and
present at the DUT connection anytime that the CAUTION HIGH VOLTAGE lamp is
ON. As in the case of many of the menu entries on the 1865 instrument, an "error
message" will be displayed for an invalid entry.