Preventive Maintenance
6-6
QIAGEN BioRobot M48 01/2003
6.2.2
Replacing the O-rings
Replacement of the O-rings of the pipettor head is required if
they look worn-out or damaged during the weekly
maintenance procedure.
1. Switch on the BioRobot M48 and open the QIAsoft M
software (Section 5.2.1, page 5-2).
2. Select “Manual Operation/Greasing O-rings”.
The pipettor head moves to the front of the worktable,
the 6 tip adapters move down and become accessible,
and the workstation door unlocks.
3. Close the QIAsoft M software and switch off the
BioRobot M48 (Section 5.2.5, page 5-5).
4. Remove the O-ring from each tip adapter using
needle-nosed pliers.
5. At the bottom of each tip adapter, wipe clean the groove
for the O-ring.
6. Ensure that the new O-rings are not turned inside out
(i.e., the O-ring should appear completely flat and not
warped).
7. Attach an O-ring at the bottom of a tip adapter.
8. Pull upward the side of the O-ring that is at the back of
the tip adapter.
9. Pull forward and upward the side of the O-ring that is at
the front of the tip adapter.
10. Check that the O-ring is fitted correctly onto the tip
adapter: the inside of the O-ring must be in complete
contact with the surface of the tip adapter.
If the upper part of the O-ring is bending outward, push
it inward using a finger or a spatula.
11. Repeat steps 7 to 10 for the remaining tip adapters.
12. Switch on the BioRobot M48 and open the QIAsoft M
software.
13. Grease the O-rings (page 6-5).
Содержание BioRobot M48
Страница 1: ...January 2003 User Manual...
Страница 48: ...General Description 3 16 QIAGEN BioRobot M48 01 2003...
Страница 92: ...Preventive Maintenance 6 8 QIAGEN BioRobot M48 01 2003...
Страница 104: ...Troubleshooting 7 12 QIAGEN BioRobot M48 01 2003...
Страница 108: ...Glossary 8 4 QIAGEN BioRobot M48 01 2003...
Страница 110: ...Appendix A A 2 QIAGEN BioRobot M48 01 2003...
Страница 116: ...Appendix B B 6 QIAGEN BioRobot M48 01 2003...