Troubleshooting
QIAGEN BioRobot M48 01/2003
7-3
6. Note the position of the pipettor head after the collision.
In addition, look for any causes of the collision.
7. Contact QIAGEN Technical Services.
Provide the following information:
- Serial number of the BioRobot M48
- Name and telephone number of the contact person
- The protocol that was run and its version number
- The type and number of samples used in the protocol
run
- The sample numbers of the block or plate in which
the collision occurred
- The position of the pipettor head after the collision,
including the position of the tip guard and the
magnet
7.2.2
Types of error log codes
If the BioRobot M48 unexpectedly develops an error, click the
QIAsoft M BIOS button to view the error log window
(Section 5.2.4, page 5-4).
Error log codes follow the format
X M S EEE
, where:
•
X
is the error level
•
M
is the module number
•
S
is the submodule number
•
EEE
is the error code
Error level
There are 3 types of error level:
• W — warning
• E — error
• F — fatal