Installation Procedures
QIAGEN BioRobot M48 01/2003
4-1
4 Installation
Procedures
The unpacking and installation of your BioRobot M48 is
carried out by a QIAGEN Instrument Service Specialist.
A member of your group who is familiar with laboratory and
computer equipment should be present during the
installation.
4.1 Requirements
Site
The BioRobot M48 must be located out of direct sunlight,
away from heat sources, and away from sources of vibration
and electrical interference. Refer to Appendix B for the
operating conditions (temperature and humidity).
Use a level workbench that is large and strong enough to
accommodate the BioRobot M48 and the computer
hardware. Refer to Appendix B for the weight and dimensions
of the BioRobot M48.
No shelves or cupboards should hang above the
BioRobot M48, since the height of the instrument increases
when the workstation door opens. To ensure that the opened
workstation door is above the user’s head, the workbench
must be at least 80 cm high.
The BioRobot M48 and the computer hardware must be
placed near to 3 properly grounded (earthed) AC power
outlets. The power lines to the equipment should be voltage
regulated and surge protected.
Computer
The operation of the BioRobot M48 requires a separate
personal computer. The computer is supplied by QIAGEN.
The minimum hardware and software specifications are
described in Appendix B.
Содержание BioRobot M48
Страница 1: ...January 2003 User Manual...
Страница 48: ...General Description 3 16 QIAGEN BioRobot M48 01 2003...
Страница 92: ...Preventive Maintenance 6 8 QIAGEN BioRobot M48 01 2003...
Страница 104: ...Troubleshooting 7 12 QIAGEN BioRobot M48 01 2003...
Страница 108: ...Glossary 8 4 QIAGEN BioRobot M48 01 2003...
Страница 110: ...Appendix A A 2 QIAGEN BioRobot M48 01 2003...
Страница 116: ...Appendix B B 6 QIAGEN BioRobot M48 01 2003...