Troubleshooting
QIAGEN BioRobot M48 01/2003
7-1
7 Troubleshooting
Refer to this section if you encounter problems while
operating the BioRobot M48. The following items are
described:
• General
troubleshooting
• Errors displayed in the QIAsoft M software
7.1 General
troubleshooting
Error Cause/Remedy
BioRobot M48 cannot
be switched on
BioRobot M48 is not receiving power.
Check that the power cord is connected to the
workstation and to the wall power outlet.
Computer cannot be
switched on
Computer is not receiving power.
Check that the power cord is connected to the
computer and to the wall power outlet.
BioRobot M48 shows
no movement when a
protocol is started
BioRobot M48 is not switched on.
Check that the BioRobot M48 is switched on.
BioRobot M48 shows
abnormal movement
when a protocol is
started
The pipettor head may have lost its home position.
In the QIAsoft M software, select “Manual
Operation/Home”.
After confirming that the pipettor head moves to
the home position, run the protocol again.
Aspirated liquid drips
from the disposable
tips
Dripping is acceptable when ethanol is being
handled. For other liquids: air is leaking from the
syringe pumps.
Grease or replace the O-rings (pages 6-5
and 6-6, respectively).
If the problem persists, contact QIAGEN
Technical Services.
Содержание BioRobot M48
Страница 1: ...January 2003 User Manual...
Страница 48: ...General Description 3 16 QIAGEN BioRobot M48 01 2003...
Страница 92: ...Preventive Maintenance 6 8 QIAGEN BioRobot M48 01 2003...
Страница 104: ...Troubleshooting 7 12 QIAGEN BioRobot M48 01 2003...
Страница 108: ...Glossary 8 4 QIAGEN BioRobot M48 01 2003...
Страница 110: ...Appendix A A 2 QIAGEN BioRobot M48 01 2003...
Страница 116: ...Appendix B B 6 QIAGEN BioRobot M48 01 2003...