Tel: (888) 545-1254 • Fax: (415) 358-4602 • Email: [email protected]
www.PrecisionRatedOptics.com
91
19.1 Common Problems and Solutions
Problem
Possible Reason
Solution
TP-P6 does not power on
Battery power is depleted
●
Charge the battery
●
Replace the battery
●
Connect the PRO TP-P6 to external power supply.
External power supply is not connected Connect PRO TP-P6 to external power supply.
External power supply is unplugged
Make sure external power supply is well connected on
both ends.
Battery dock cover is not closed
properly
Close the battery dock cover properly
Black Screen
Screen backlight is off, but power
indicator is on.
Press any key to exit “Power Saving” mode.
Battery depleted, the PRO TP-P6 is off
and the power indicator is off.
●
Charge the battery
●
Replace the battery
●
Connect the PRO TP-P6 to external power supply.
External keypad no response
External keypad is not recognized.
●
Unplug the keypad and re-plug
●
Power off PRO TP-P6, plug keypad and
power on PRO TP-P6 again.
External mouse no response
External mouse is not recognized.
●
Unplug the mouse and re-plug
●
Power off PRO TP-P6, plug in the mouse and
power on the PRO TP-P6 again.
Screen display blurs
Brightness is not set properly
Adjust brightness
Application no response
Problem error
Restart PRO TP-P6 and run the application again.
Operation time is short
Battery run out
Change the battery
Battery is not fully charged
●
Charge the battery
●
Re-calibrate the battery
Touchscreen is not responding
properly with stylus
Touch screen needs calibration
Connect external USB mouse to calibrate the stylus.
PRO TP-P6 indicates low
battery after battery is fully
charged
Battery is not fully charged.
Re-calibrate the battery
19.2 Trace Measurement Problems and Solutions
Problem
Possible Reason
Solution
Expected events are not shown
in the trace
Events are too close to each other.
●
Try again with short pulsewidth
●
Try testing from the other end of the fiber
SNR is poor
Try again with wide pulsewidth or increase averaging time.
Wrong configuration
Check OTDR parameters, like Reflection Threshold,
Non-Reflection Threshold and End Threshold. Re-analyze
or test again.
Special events appear at the
end of the trace
Wrong configuration, e.g. wrong end
threshold setting.
Adjust configuration and test again.
The fiber under test is too long to test
accurately.
Select proper range and test again.
19 Troubleshooting
Содержание TP-P6
Страница 1: ...TP P6 Manual Multifunctional Test Platform...