Tel: (888) 545-1254 • Fax: (415) 358-4602 • Email: [email protected]
www.PrecisionRatedOptics.com
56
11.14.1 Display the Event on the trace, and locating the event in the Event Table
At any time, while scrolling through the event table, the user can check the relative info about all events
tested on a particular trace. When an event is selected, a
Marker
will be placed at that event.
Note:
●
When the trace is under magnification, certain Markers may not be displayed in the current window.
●
It is adviseable to first view the entire trace, locate the events, then zoom into specific events to
analyze specific details.
11.14.2 Add an Event
The user can manually insert events into the event table at any time. For example, when there is a splice
point on some given position, but is hidden due to excessive noise, or the splice loss is lower than the
minimum set value of a non-reflection detection threshold, then adding an event can be useful. Adding an
event manually will add an event numberto the insertion point, but will not otherwise alter the trace.
Note:
When reanalyzing a trace, any manually inserted events will be automatically deleted from the trace.
To add an event:
1. Press [add] while inside the {Events} tab.
Fig 11.3 Adding an Event
11 Trace Analysis and Event Operation
Содержание TP-P6
Страница 1: ...TP P6 Manual Multifunctional Test Platform...