18.9 Battery Calibration
After the battery either sits unused for a long period of time, or has been used frequently, the battery
power level monitor may not be accurate. In some cases, for example, the battery can indicate that it is
at least partially charged, yet the unit shuts down automatically due to a dead battery. If this occurs, the
user will need to recalibrate the battery.
Procedure
:
1. Discharge the battery completely, leaving the unit on until the unit shuts down automatically.
2. Charge the battery until fully charged.
18.10 Change Battery
1. Power-off the PRO TP-P6 and unplug the power adapter.
2. Open the battery dock cover.
3. Change the battery.
4. Re-attach the battery dock cover.
Note:
●
Please Note: Always turn off the TP-P6 before performing a battery replacement.
●
Only authorized service personnel can change the internal clock battery.
18.11 Calibration Requirements
Calibration of the PRO TP-P6 is recommended every two years. Please contact your nearest PRO
representative, or contact our Customer Service Center for instructions on sending in the TP-P6 for
proper calibration.
18.12 Shipment
The PRO TP-P6 should always be shipped in its original packing materials, if possible. Otherwise, the
TP-P6 should be professionally packed for safe shipment. Precision Rated Optics is not liable for damage
that may occur during shipping, due to incorrect or inadequate packing procedures.
Please note the following points when shipping the PRO TP-P6:
●
Use original packing materials, if possible.
●
Avoid excessive temperature and humidity changes.
●
Avoid direct exposure to sunlight.
●
Avoid shock and vibration.