P
AGE
33
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
My Setup:
When you select “My Setup”, you will be able to view the following screens. You can change the settings in
the Setup menu. These screens will only display if they are available and enabled on the device.
Icon
Text
Description
Ramp
This displays the ramp starting pressure. You can increase or decrease
the ramp starting pressure in 0.5 cm H
2
O increments.
Ramp Time
When you set the Ramp time, the device increases the pressure from
the value set on the Ramp screen to the therapy pressure setting over
the length of time specified here.
Rise Time
Rise time is the time it takes for the device to change from EPAP to
IPAP. If rise time is prescribed for you, you can adjust the rise time from
1 to 6 to find the setting that provides you with the most comfort. A setting
of 1 is the fastest rise time, while 6 is the slowest.
Flex
When in S mode and Bi-Flex is enabled, you can adjust the Bi-Flex
setting. This allows you to adjust the level of air pressure relief that you
feel when you exhale during therapy. Your home care provider can
enable or disable this feature. When your provider enables Flex, a level
will already be set for you on the device. You can increase or decrease
the setting from 1 to 3. The setting of 1 provides a small amount of
pressure relief, with higher numbers providing additional relief.
Note: If a lock icon is displayed on this screen, it indicates that your
provider has locked this setting and you cannot change it
Humidification
This displays the Humidification Mode being used. You can choose
between Fixed or Adaptive Humidification. If a heated tube is being
used, the device will automatically switch to Heated Tube Humidification
Mode. A “lock” symbol will appear next to the mode setting indicating
that so long as the heated tube is attached to the device, this mode
cannot be changed. However, the heater plate and tube temperature
settings can still be adjusted on the device Therapy screen as normal.