Alert
Icon
Priority
Description
User Action
Patient Message
Notification
Message from your
Provider.
Press Control Dial to
acknowledge and clear the
message.
Change Rejected
Alert
A prescription or
settings change was
rejected.
N/A
Humidification Error.
Contact support if the
problem persists.
Status
Humidifier error (only
when humidifier is pre-
sent)
Humidifier heater
plate error or
humidifier not
properly connected
to therapy device.
Turn off device and discon-
nect from power. Detach the
humidifier, visually check that
electrical contacts are clear,
then reconnect humidifier and
power cord.
Heated Tube Error.
Contact support if the
problem persists
Status
Heated tube error
(only when heated
tube is present).
Heated tube may be
overheated or dam-
aged.
Turn off device. Detach
heated tube from
humidifier, make sure that
tube is not covered or
obstructed, and then
reattach to humidifier. If
alert continues,
troubleshoot for possible
failure.
The attached power
supply does not support
humidification.
Alert
Indicates that the
attached power
supply is not
capable of
supporting
humidification or
heated tube.
Switch to a Philips
Respironics DreamStation
power supply that is
capable of supporting
humidification. Or operate
therapy device without
humidifier.
Activate MODE?
Notification
Indicates an incompati-
ble power supply is at-
tached. Incompatible
power supply, or
power cord is not fully
inserted into device’s
power inlet.
Confirm power cord is fully
inserted into device’s power
inlet. Confirm a compatible
Philips Respironics power
supply is attached. Switch to
compatible power supply if
needed.