![PHILIPS Respironics DreamStation BiPAP AVAPS Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/philips-respironics/dreamstation-bipap-avaps/dreamstation-bipap-avaps_service-and-technical-reference-manual_1550628036.webp)
P
AGE
36
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
Provider Settings
Therapy Settings:
Choosing this screen will take you to a sub-menu where you can adjust the device therapy modes and pres-
sure settings. These settings are described here.
Note:
Not all settings shown here will display on the device. The display will vary based on therapy device
model and device settings.
Icon
Text
Description
CPAP
S
S/T
T
PC
Mode
This setting allows you to select a therapy mode. The default
setting is S/T. The available modes are:
BiPAP S/T:
CPAP, S, or S/T
BiPAP AVAPS:
CPAP, S, S/T, T, or PC
Note: If the therapy mode is changed while the blower is turned
on, a confirmation pop-up screen appears. Select Yes if you
would like to activate the selected mode.
AVAPS
AVAPS
This screen allows you enable or disable AVAPS. Select On to
enable AVAPS and Off to disable.
AAM
AAM
This setting allows you to enable or disable Automated Airway
Management (AAM). Select On to enable AAM and Off to dis-
able.
Max Pressure
This setting displays the current maximum pressure setting.
You can adjust the setting as follows:
25 cm device: 4 to 25 cmH
2
O in 0.5 increments
30 cm device: 4 to 30 cmH
2
O in 0.5 increments
EPAP Min
This setting allows you to modify the Minimum EPAP setting.
This setting will be the minimum level of pressure applied dur-
ing the expiratory breath phase. You may adjust the setting as
follows:
25 cm device: 4 cmH
2
O up to the lesser value of either maxi-
mum pressure or 21 cmH
2
O
30 cm device: 4 cmH
2
O up to the lesser value of either maxi-
mum pressure or 25 cmH
2