P
AGE
80
1127179,
V
ER
.
02
D
REAM
S
TATION
B
I
PAP
S/T
AND
AVAPS
S
ERVICE
&
T
ECHNICAL
R
EFERENCE
M
ANUAL
6.7 Device Alarms
There are two types of alarms:
•
High priority - Require immediate response by the operator.
•
Medium priority - Require prompt response by the operator.
When an alarm condition occurs:
•
The audio pause icon appears and the control dial LED indicator lights.
•
The audible alarm sounds.
•
A message appears on the screen describing the type of alarm.
6.8 Device Alerts
Alerts are informational and confirmation messages that notify you of conditions that require attention but are
not alarm conditions. Most alerts do not appear while therapy is being delivered. The different types of alerts
are:
•
Status - Message appears on the screen.
•
Notification - Message appears on the screen and Therapy button LED flashes.
•
Alert - Message appears on the screen, Therapy button LED flashes, and audible sound
beeps once while alert is displayed.
6.9 Alarm and Alert LED Indicators
Red flashing alarm LED
When a high priority alarm is de-
tected, the audio pause icon (1)
and control dial LED (2) flash
red.
Yellow flashing alarm LED
When a medium priority alarm is
detected, the audio pause icon
(1) and control dial LED (2) flash
yellow.
Flashing Therapy button LED
When an alert or notification
message appears on the screen,
the Therapy button LED flashes.
If the alert occurs while therapy
is being delivered, the alert or
notification will appear on the
screen, but the LED will not
flash.