To install the Blower, Blower Box Assembly and Rear Panel:
1. Position the Blower inside the Blower Box Assembly so that the Blower Outlet Seal and Blower
Isolators are aligned in their grooves.
CAUTION
•
The Blower with the Isolators attached should be rested inside the Blower Box. Do not
press down on the Blower after installation.
2. Insert Grommet into slot.
3. Set Blower wire length to approximately 9.5 in from the outside of the Blower Box Assembly.
Blower Outlet Seal
Blower Isolators
Ensure wires are
routed as shown
Grommet
Wire length approx. 9.5