1.4    Service Notice 

The service technician should have a good working knowledge and understanding of the principles of opera-
tion and repair of electro-mechanical sleep therapy devices. By using the most current version of the service 
manual (found on my.respironics.com) and the latest testing software, all repairs and testing can be per-
formed. If service training is desired, contact the Philips Respironics service location in your area to schedule 
training. 

1.5  Service Training 

Philips Respironics offers service training for the devices. Training includes complete disassembly of the de-
vice, troubleshooting sub-assemblies and components, and necessary safety testing. For more information, 
log onto my.respironics.com, and download the Service Training Schedule brochure from the Service Soft-
ware and Documentation link. 

Содержание DreamStation BiPAP AVAPS

Страница 1: ...PAGE 1 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL DreamStation BiPAP S T and AVAPS Service and Techincal Reference Manual...

Страница 2: ...C Power to the Device 16 5 2 Supplying AC Power to the Device 17 5 3 Connecting the Tubing to the PAP Device 18 5 4 Connecting the Humidifier to the PAP Device 19 5 5 Connecting the Tubing to the Humi...

Страница 3: ...the SD Card 49 Chapter 6 Troubleshooting and Error Codes 50 6 1 Introduction 50 6 2 Bench Checkout 50 PAP Device 50 Humidifier 50 6 3 System Verification 51 Pressure Verification 51 Heated Humidifier...

Страница 4: ...e and SD Flip Doors 94 Replacing the SD Card 95 Replacing the Upper Enclosure Keypad 95 Replacing the UI Panel 97 Replacing the PCA 98 Replacing the Flow and Pressure Sensor Seals 101 Replacing the Bl...

Страница 5: ...Replacing the Lifting Tray 126 Replacing the Bottom Cover 129 Replacing the Heat Shield 130 Replacing the Heater Plate O Ring 131 Replacing the Bottom Flip Lid Assembly 132 Chapter 9 Testing 137 9 1 R...

Страница 6: ...PAGE 6 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 2017 Koninklijke Philips N V All rights reserved...

Страница 7: ...C or W DreamStation BiPAP S T 30 yyX1030Szz C or W DreamStation BiPAP S T 30 w Humidifier and Standard Tube yyX1030Hzz C or W DreamStation BiPAP S T 30 w Humidifier and Heated Tube yyX1030Tzz C or W D...

Страница 8: ...des increased pressure relief during the expiratory phase of breathing United States The BiPAP S T and AVAPS devices are intended to provide non invasive ventilatory support to treat adult and pediatr...

Страница 9: ...irway The AAM feature adjusts the EPAP level between the minimum EPAP Min and maximum EPAP Max settings The IPAP level is controlled by the pressure support PS setting AVAPS BiPAP AVAPS device only If...

Страница 10: ...ramp period If ramp is activated with AVAPS and Automated Airway Management AAM disabled it will reduce EPAP and IPAP pressures to Ramp Start pressure and Ramp start pressure plus a Delta and ramp up...

Страница 11: ...me to find the most comfortable setting for the patient Rise time cannot be adjusted when Bi Flex is enabled Digital Auto Trak Digital Auto Trak is an important ventilation feature due to its ability...

Страница 12: ...is comprised of the following components Heated Humidifier The heated humidifier is the primary source of humidification Humidification is controlled by adjusting the temperature of the heater plate T...

Страница 13: ...com and the latest testing software all repairs and testing can be per formed If service training is desired contact the Philips Respironics service location in your area to schedule training 1 5 Ser...

Страница 14: ...n equipment or property damage and important information that must be considered when performing service and testing procedures on the device WARNING Warnings indicate the possibility of injury to peo...

Страница 15: ...he power cord from the wall outlet before cleaning the device DO NOT immerse the device in any fluids Perform the following steps to clean and disinfect the device and humidifier prior to servicing th...

Страница 16: ...rs before beginning setup Do not use extension cords with this device NOTE Please refer to the Clinical Manual for additional information 5 1 Supplying DC Power to the Device A Philips Respironics DC...

Страница 17: ...ed end of the AC power cord into an electrical outlet that is not controlled by a wall switch 3 Plug the power supply cord s connector into the power inlet on the side of the device 4 Verify that the...

Страница 18: ...on the back of the therapy device Line up the connector 1 at the top of the heated tube to the top of the air outlet port on the back of the device 2 Press the tubing into place over the air outlet po...

Страница 19: ...ubing to prevent contamination 5 4 Connecting the Humidifier to the PAP Device CAUTION Do not move the humidifier while the water tank has water in it 1 Place the therapy device and heated humidifier...

Страница 20: ...e connector 1 at the top of the heated tube to the top of the air outlet port 2 on the humidifier 2 Press the tubing into place over the air outlet port until the tabs on the side of the tube click in...

Страница 21: ...wn instead of a heated tube simply pull the tubing away from the outlet port 5 7 Disconnecting the Devices CAUTION To avoid spilling do not disconnect the humidifier from the therapy device with water...

Страница 22: ...laced To install or reseat your humidifier lid seal fully open the humidifier lid Position the seal 1 against the inside of the lid so the seal s center hole aligns with the humidifier outlet port Con...

Страница 23: ...ace at all times when the device is operating The ultra fine filter is rec ommended for people who are sensitive to tobacco smoke or other small particles The reusable blue filter is supplied with the...

Страница 24: ...ss door and swing open If replacing pull out the old filter assembly 2 If applicable place a clean reusable blue pollen filter 1 on top of a new optional disposable light blue ultra fine filter 2 and...

Страница 25: ...ur local time zone If you choose to skip this initial time setting the time can always be adjusted under the My Setup menu Note This time setting is not displayed as a clock function on the device It...

Страница 26: ...To adjust a setting 1 Rotate the control dial to your desired setting 2 Press the control dial to select that setting 3 Rotate the control dial to change the setting 4 Press the control dial again to...

Страница 27: ...dification Settings While the device is delivering therapy you can adjust Tube Temperature or Humidifier Settings Rotate the control dial to choose either setting Press and rotate the dial to change t...

Страница 28: ...ss the control dial to save the setting Note If you are using the humidifier without the heated tube simply just rotate the control dial in either direc tion to change the humidifier setting Ramp Feat...

Страница 29: ...Info you will be able to view the following screens You cannot change settings in the Info menu These screens are only for reference Note Additional icons may appear if optional accessories are being...

Страница 30: ...eriodic breathing indicated her contact your home care provider for assistance This screen only displays if your home care provider has enabled it 90 Pressure 90 Pressure This screen displays the nigh...

Страница 31: ...oaded to your device Phone In This screen displays the total therapy hours for the device the total blower hours the total number of days used when the sessions were greater than 4 hours and a complia...

Страница 32: ...te an upload a second time or contact your home care provider if the issue persists This screen is locked if modem is off Device Info This screen displays your therapy device information serial number...

Страница 33: ...west Flex When in S mode and Bi Flex is enabled you can adjust the Bi Flex setting This allows you to adjust the level of air pressure relief that you feel when you exhale during therapy Your home car...

Страница 34: ...this setting and you cannot change it Language This feature allows you to choose which language to display on the interface You can also turn off 0 text mode which means the device will display the Ic...

Страница 35: ...e control dial Rotate the control dial in either direc tion to scroll through the menu options on the display screen To adjust a setting 1 Rotate the control dial to your desired menu option 2 Press t...

Страница 36: ...on pop up screen appears Select Yes if you would like to activate the selected mode AVAPS AVAPS This screen allows you enable or disable AVAPS Select On to enable AVAPS and Off to disable AAM AAM This...

Страница 37: ...that is per mitted between IPAP and EPAP You may adjust the setting from the minimum pressure support setting to the difference between the maximum pressure setting minus the minimum EPAP setting IPA...

Страница 38: ...Inspiratory Time setting You may adjust the setting from 0 5 to 3 0 seconds in 0 1 in crements This setting only displays if PS max is greater than zero and BPM is not set to Off or Auto Vt Vt This sc...

Страница 39: ...1 0 increments The alarm will sound when the calcu lated minute ventilation is less than or equal to the specified setting The default is 0 Low Tidal Vol ume You can enable or disable this alarm by s...

Страница 40: ...nditions in the room and is designed to not allow condensation to occur in the tube Humidifier This setting allows you to choose the desired humidity setting for the humidifier 0 1 2 3 4 or 5 Tube Tem...

Страница 41: ...ing that you are using with the device You can choose either 22 for the Philips Respironics 22 mm tubing or 15 for the Philips Respironics 15 mm tubing When using Heated Tubing the device will automat...

Страница 42: ...on the interface Clear Default Reminders This setting turns off the default patient reminders that are enabled in the therapy device from the factory Note This does not turn off additional reminders t...

Страница 43: ...e unlocked icon appears on the screen while in this mode Info Screens Choosing this screen will take you to a sub menu where you can view information on patient usage The Info screens are described in...

Страница 44: ...ERA algorithm monitors for a sequence of breaths that exhibit both a subtle reduction in airflow and progressive flow limitation If this breath sequence is terminated by a sudden increase in airflow a...

Страница 45: ...ilable networks within range of your device Press the control dial to select your network 3 If your network is secure a Passkey pop up screen appears Enter your network password and then select the gr...

Страница 46: ...Follow the steps below to manually pair to your mobile phone or tablet 1 With your therapy device powered up and the blower off initiate Bluetooth Setup from the DreamMapper mobile app 2 If you need t...

Страница 47: ...nd the screen will either display a green checkmark or a red X The green checkmark indicates that the leak found allows for optimal performance of the device The red X indicates that the leak may affe...

Страница 48: ...shooting dashboard 5 19Accessories There are several accessories available for the DreamStation system such as a Humidifier Cellular Modem Wi Fi Accessory or a Link Module When using optional accessor...

Страница 49: ...32 or DB9 port to allow remote control of the DreamStation BiPAP autoSV device by a personal computer Oximeter The DreamStation oximetry system when part of the DreamStation BiPAP autoSV device measur...

Страница 50: ...ice pressure by using a manometer 7 Listen to the device for noisy operation or loose components 8 Verify the alarms function properly refer to section 6 3 3 9 Refer to section 6 4 to retrieve the dev...

Страница 51: ...l Manometer REF 302227 or equivalent with the following minimum specifications o 0 30 cm H2O or better o 0 3 cm H2O accuracy o 0 1 cm H2O resolution To verify the pressure on a BiPAP S T or AVAPS devi...

Страница 52: ...e if the DreamStation Humidifier is working properly The following steps should be followed to confirm the performance of the humidifier WARNINGS It is important to follow the exact steps below when p...

Страница 53: ...t You must verify that the Patient Disconnect Alarm operates properly with the prescribed patient pressures and circuit 1 Set the Patient Disconnect Alarm setting to 15 seconds 2 Exit to the Home scre...

Страница 54: ...therapy 2 Simulate 6 breaths by alternately occluding and opening the outlet port for 2 seconds each 3 With the therapy still on set the Low Minute Ventilation Alarm setting 10 0 LPM 4 Simulate 1 or 2...

Страница 55: ...ours Humidifier Verification Test Result circle one Humidification test heater plate operation Pass Fail System and Alarm Verification Test Result Tolerance Result circle one CPAP 4 cmH2O 1 cmH2O Pass...

Страница 56: ...ite The Service Center Tools Suite will provide you the necessary tools to view the device s error event log along with additional functions necessary to service the device To download the software yo...

Страница 57: ...Click on the Download button adjacent to the Service Center Tools Suite v3 8 or greater 3 Save the Service Center Tools Installer to your PC default directory is recommended 4 To use the Service Cent...

Страница 58: ...on allows you to set the session ID on the device This should only be executed when the Therapy PCA is replaced on a device Read Time Meters This function allows you to read the therapy and blower hou...

Страница 59: ...nt Otherwise the therapy hours should remain on the de vice If the PCA is NOT replaced during device servicing the blower hours should be cleared only if the Blower is replaced Otherwise the blower ho...

Страница 60: ...hould appear indicating the Session ID has been set 5 Select the OK button 6 Select the New Device button if you are setting sessions on multiple devices otherwise select the Close button to exit the...

Страница 61: ...ted by the system to a STOP level error ABORT The error information is recorded in NVRAM The fifth occurrence of an ABORT level error within a 24 hour period while power is maintained will be promoted...

Страница 62: ...ERR_SOFTWARE REBOOT CPU 1 If error was NOT Last Stop Error clear error log and test 2 If error was Last Stop Error replace PCA and test E 8 ERR_CRC_FAILED STOP CPU 1 Replace PCA and retest E 10 ERR_W...

Страница 63: ...urs replace blower and retest 3 If 2nd retest fails replace PCA E 21 ERR_MOTOR_VBUS_HI GH STOP PCA 1 Replace PCA and retest E 22 ERR_MOTOR_FLUX_MA GNITUDE REBOOT 1 Motor connection 2 Blower Box 3 PCA...

Страница 64: ...rs replace blower and retest 3 If 2nd retest fails replace PCA E 29 ERR_VBUS_GAIN_ZERO REBOOT PCA 1 If E9 was not recorded clear error log and test 2 If E9 was recorded replace PCA and test E 30 ERR_M...

Страница 65: ...Stop Error replace PCA and test E 48 ERR_NVRAM_MAX_RET RIES_EXCEEDED REBOOT PCA 1 If error was NOT Last Stop Error clear error log and test 2 If error was Last Stop Error replace PCA and test E 50 ERR...

Страница 66: ...ET_STOP STOP PCA 1 Replace PCA and test E 74 ERR_PSENS_NO_CALLB ACK REBOOT PCA 1 If error was NOT Last Stop Error clear error log and test 2 If error was Last Stop Error replace PCA and test E 82 ERR_...

Страница 67: ...r Plate 1 Clear error log and test 2 Test humidifier separately E 101 ERR_HUMID_MAX_TEM P CONTINUE Heater Plate 1 Clear error log and test 2 Test humidifier separately E 105 ERR_HUMID_AMBIENT_ COMM CO...

Страница 68: ..._RAM REBOOT PCA 1 If error was NOT Last Stop Error clear error log and test 2 If error was Last Stop Error replace PCA and test E 150 ERR_CPU_NMI ABORT PCA 1 Replace PCA and test E 151 ERR_CPU_HARD_FA...

Страница 69: ...RR_NVRAM_MUTEX_N OT_AVAILABLE REBOOT PCA 1 If error was NOT Last Stop Error clear error log and test 2 If error was Last Stop Error replace PCA and test E 182 ERR_NVRAM_WRITE_FA ILED REBOOT PCA 1 If e...

Страница 70: ..._HOST_NO_D EVICE CONTINUE Accessory Module 1 Remove and reinsert the accessory module 2 If error continues replace accessory module if pulse oximetry is required E 200 ERR_RTOS_ABORT_ER ROR STOP PCA 1...

Страница 71: ...st 2 If error was Last Stop Error replace PCA and test E 222 ERR_MOTOR_DIV_ZERO CONTINUE N A 1 Clear error log and retest E 223 ERR_LOW_PRESSURE_ REGULATION CONTINUE N A 1 Load latest software version...

Страница 72: ...unit remove that first and retest 2 Load latest software version if not up to date 3 Clear air path including filter and test 4 If test fails replace PCA and retest E 227 ERR_LOW_MOTOR_RP M CONTINUE 1...

Страница 73: ...fails or the error reoccurs replace the PCA and retest E 260 ERR_CIRCUIT_COMP_M ODEL_DIV_ZERO_1 CONTINUE N A 1 Clear error log and test E 261 ERR_CIRCUIT_COMP_M ODEL_DIV_ZERO_2 CONTINUE N A 1 Clear er...

Страница 74: ...d in the whole way If returned check to see if the power supply is correct Make sure power cord is pushed in the whole way No Power All connections not made Power Supply not functioning PCA Failure Da...

Страница 75: ...gned on seals Make sure enclosures are secure Check for any component damage Check for loose components Check Blower wire length outside of Blower Box should measure approx 9 5 inches Replace any fail...

Страница 76: ...Tube settings Unit Shuts Off Auto Off feature is enabled and high leak exists for a duration that triggers Auto Off System Issue Power Supply faulty Examine Auto Off setting in provider mode Check Pow...

Страница 77: ...ections Unplug the device and reapply power Relocate the device to an area with lower EMI emissions cellular phones computers TVs etc Check UI connections Replace UI Panel if faulty Back Light Does No...

Страница 78: ...PAP Electrical issue with Humidifier 80W Power Supply not being used User perception Ensure tube setting is anything but 0 and verify the Tube is warming Ensure you use an 80W Power Supply when testi...

Страница 79: ...een PAP and Modem not functioning Verify the PAP recognizes the Modem in the My Provider menu Attempt to make a call and verify it was successful SD Card Error SD Card corrupted Problem with PCA Eject...

Страница 80: ...e therapy is being delivered The different types of alerts are Status Message appears on the screen Notification Message appears on the screen and Therapy button LED flashes Alert Message appears on t...

Страница 81: ...er failure When a power failure occurs a series of beeps sounds in a 1 beep pat tern repeating one second on then one second off High priority When a high priority alarm is active a series of beeps so...

Страница 82: ...alarm Press the control dial again to remove the alarm screen from the display Resetting the alarm allows you to return to the previous screen If multiple alarms occur during the same period of time t...

Страница 83: ...nowledge and remove the alarm from the screen High Operates This alarm occurs when the patient has not triggered a breath within the time speci fied in the apnea alarm set ting The device continues to...

Страница 84: ...inop erative Shuts down Press the control dial to si lence the alarm Disconnect the device from power Device inop erative Shuts down Check device air inlet is not obstructed Check air filter s are in...

Страница 85: ...arm from the screen Remove your mask Check for the following dirty inlet filters blocked air intake excessive leak in the patient circuit High Operates if the alarm continues for 10 seconds the alarm...

Страница 86: ...Action Medium Operates Press the control dial again to acknowledge and remove the alarm from the screen Re move your mask Remove power from the device Con firm a compatible Philips Respironics power s...

Страница 87: ...write underway No action needed Change Accepted Status Confirms acceptance of prescription change or device upgrade No action needed Oximetry Good Connec tion icon only Status Displays on the ther ap...

Страница 88: ...ate source of ventilation SD Card Removed Notification or Alert 2 Indicates SD card has been removed from therapy device and not reinserted before the start of the current therapy session Reinsert SD...

Страница 89: ...tatus Heated tube error only when heated tube is present Heated tube may be overheated or dam aged Turn off device Detach heated tube from humidifier make sure that tube is not covered or obstructed a...

Страница 90: ...y inaccessible due to data communication No action needed Three Night Summary N A Status Displays last 3 nights of hourly use Press control dial to acknowledge and clear the screen Otherwise message t...

Страница 91: ...fter any repairs are made Refer to Chapter 9 for details on testing the device WARNING To prevent electrical shock disconnect the electrical supply before attempting to make any repairs to these devic...

Страница 92: ...y BiPAP AVAPS 1130520 UI Knob 1130437 PCA BiPAP S T 25 IN DE CN 1130523 PCA BiPAP S T 30 US FR BR 1130734 PCA BiPAP S T 30 IN DE CN 1130737 PCA BiPAP AVAPS 25 IN 1130524 PCA BiPAP AVAPS 30 US 1130758...

Страница 93: ...1132247 Warning Label AU FR 1132278 Warning Label CA 1125963 Power Supply 80W 1118499 Travel Power Supply DOM 65W 1120136 Travel Power Supply EU 65W 1127194 Travel Power Supply CN 65W 1127195 Travel...

Страница 94: ...oors Kit Included in Kit Tools Required Accessory Module Flip Door Accessory Module Flip Door None SD Cover Flip Door Accessory Cover Flip Door To remove the Flip Doors 1 Squeeze the Flip Door standof...

Страница 95: ...it locks into place Replacing the Upper Enclosure Keypad Kit Included in Kit Tools Required Upper Enclosure Assembly Upper Enclosure Keypad 4 x 1 2 screws qty 2 Torx Screwdriver T10 is recommended To...

Страница 96: ...nto place 2 Secure the Upper Enclosure to the Bottom Enclosure using the two 4 x 1 2 torque to 6 IN LB 3 Install the Keypad on the Upper Enclosure as shown in Figure 6 3 above and verify it is fully s...

Страница 97: ...fting up on the ribbon 3 Detach the main UI Panel ribbon from the PCA by lifting up on the connector latch and pulling the rib bon away from the connector To install the UI Panel 1 Align the UI Panel...

Страница 98: ...Required PCA PCA 4 x 1 2 screws qty 2 Torx Screwdriver T10 is recommended To remove the PCA 1 Remove all components as instructed in the previous sections 2 Remove the two 4 x 1 2 screws that secure t...

Страница 99: ...A onto the Blower Cap standoff and screw bosses ensuring to position the Humidifier Connector located on the PCA into the groove of the Rear Panel and positioning the Flow and Pres sure Sensors into t...

Страница 100: ...NUAL CAUTION The PCA s Flow and Pressure Sensors must be in proper alignment with the Seals Otherwise the device will not operate properly Align humidifier connector into Rear Panel groove Align screw...

Страница 101: ...e Seals 1 Remove all components as instructed in the previous sections 2 Lift the Seals away from the Blower Box Assembly Note The Pressure Sensor Seal may remain attached to the Pressure Sensor on th...

Страница 102: ...ous sections 2 Remove the four 4 x 1 2 screws holding the Blower Cap to the Blower Box Assembly 3 Using a small flathead screwdriver carefully pry the Blower Cap off of the Blower Box Assembly around...

Страница 103: ...p with the screw holes on the Blower Box Assembly verifying the Blower Cap is fully seated 2 Secure the Blower Cap to the Blower Box Assembly using the four 4 x 1 2 screws torque to 2 5 IN LB 3 Assemb...

Страница 104: ...Assembly and Rear Panel 1 Remove all components as instructed in the previous sections 2 Remove the two 4 x 1 2 screws holding the Blower Box Assembly to the Bottom Enclosure 3 Lift the back end of t...

Страница 105: ...hing the ferrite straight down 7 Unwrap the wire around the Blower Box Assembly 8 Pull the wire grommet out of the Blower Box Assembly groove 9 Carefully lift the Blower straight out of the Blower Box...

Страница 106: ...l and Blower Isolators are aligned in their grooves CAUTION The Blower with the Isolators attached should be rested inside the Blower Box Do not press down on the Blower after installation 2 Insert Gr...

Страница 107: ...7179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 4 Verify the Blower Outlet Seal is properly seated GOOD Seated BAD Not Seated 5 Route the Blower wires as shown below Se...

Страница 108: ...108 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 6 Snap DC Cable ferrite into groove on Rear Panel Slide ferrite up to the top of the groove Snap ferrite into gr...

Страница 109: ...PAGE 109 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 7 Attach Rear Panel onto Blower Box Assembly Note Ensure O Ring is properly seated GOOD BAD GOOD BAD...

Страница 110: ...he device as instructed in the previous sections Replacing the Blower Outlet Seal and Blower Isolators Kit Included in Kit Tools Required Blower Outlet Seal Blower Outlet Seal Torx Screwdriver T10 is...

Страница 111: ...Seal and Blower Isolators on the Blower and verify they are fully seated 2 Assemble the remainder of the device as instructed in the previous sections GOOD CAUTION The Isolators are prone to falling o...

Страница 112: ...sections Note It is not required to remove the Blower Upper Cap Blower or Blower Isolators these can be left assembled and lifted from the Bottom Enclosure as one assembly 2 Turn the Bottom Enclosure...

Страница 113: ...nents as instructed in the previous sections it is not required to remove the Blower Upper Cap Blower or Blower Isolators these can be left assembled and lifted from the Bottom En closure as one assem...

Страница 114: ...tion on the serial model number labels is required to be duplicated in order to maintain proper traceability Refer below for details on reprinting the labels NOTE The new label MUST include the same M...

Страница 115: ...supported printers and fees please refer to http www loft ware com support tech_printers cfm For software support visit their website at www loftware com Label Printing Options There are three differe...

Страница 116: ...first be executed refer to Chap ter 6 for troubleshooting procedures IMPORTANT NOTE The device must be tested after any repairs are made Refer to Chapter 9 for details on testing the device WARNING T...

Страница 117: ...l 1120613 Flip Lid Bottom Assembly 1120873 Flip Lid Seal 1120617 Flip Lid Latch 1121569 Back Panel 1120614 Lifting Tray 1121570 Bottom Cover 1120616 Heat Shield 1122522 Heater Plate O Ring 1120669 Wir...

Страница 118: ...s Replacing the Water Tank Assembly Kit Included in Kit Tools Required Tank Assembly Tank Assembly None To remove the Water Chamber Assembly 1 Pull back on the latch located on the Bottom Assembly and...

Страница 119: ...cluded in Kit Tools Required Flip Lid Seal Flip Lid Seal None Dry Box Seal Dry Box Seal To remove the Seals 1 Remove the Water Tank as instructed in the previous section not required for Flip Lid Seal...

Страница 120: ...s one way Be sure to press the Seal the whole way around to ensure it is seated properly Replacing the Dry Box Assembly Kit Included in Kit Tools Required Dry Box Assembly Dry Box Assembly None To rem...

Страница 121: ...tructed in the previous sections Replacing the Wire Guard Kit Included in Kit Tools Required Wire Guard Wire Guard Flathead Screwdriver To remove the Wire Guard 1 Remove the Water Tank and Flip Lid Se...

Страница 122: ...operly seated by wiping your finger across Wire Gard in location shown in the illus tration below If Wire Guard is not flush with the Flip Lid surface remove Wire Guard re seat wires and reinstall Wir...

Страница 123: ...PAGE 123 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL...

Страница 124: ...ack Panel Assembly 1 Remove the Water Chamber Assembly refer to previous corresponding section 2 Pull the Back Panel away from the Humidifier while lifting up on the top to release the Back Panel tab...

Страница 125: ...ve the Flip Lid Latch 1 Remove the Back Panel Assembly and open the Flip Lid refer to previous corresponding section 2 Push up on the 2 Latch tabs and pull the Latch out of the Humidifier Assembly To...

Страница 126: ...VAPS SERVICE TECHNICAL REFERENCE MANUAL Replacing the Lifting Tray Kit Included in Kit Tools Required Lifting Tray Lifting Tray None To remove the Lifting Tray 1 Push out one side of the Lifting Tray...

Страница 127: ...ifting Tray 1 Place one end of the Lifting Tray back into the cut out groove and then repeat on other side Note The Tray will only fit one way into the groove Refer to illustration below 2 Once both e...

Страница 128: ...PAGE 128 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL Pull back Tray and align into groove...

Страница 129: ...ve the four 6 x 1 4 screws securing the Bottom Cover to the Bottom Assembly 3 Lift the Bottom Cover from the Humidifier To install the Bottom Cover 1 Align the Bottom Cover onto the Bottom Assembly 2...

Страница 130: ...at Shield Heat Shield Torx Screwdriver T15 is recommended To remove the Heat Shield 1 Remove the Bottom Cover as instructed in the previous section 2 Remove the Heat Shield from the Bottom Cover To in...

Страница 131: ...s section 2 Lift up the Heater Plate 3 Pull the O Ring from the Bottom Assembly To install the Heater Plate O Ring 1 Place the O Ring into the groove on the Bottom Assembly as shown in the above 2 Pla...

Страница 132: ...Assembly Bottom Flip Lid Assembly Dry Box Heater Plate Heater Plate O Ring Heat Shield Bottom Cover Back Panel 6 x 1 4 screws qty 4 Torx Screwdriver T10 and T15 is recommended ISO Port ISO Port None...

Страница 133: ...g sections 2 Install Back Panel Assembly Flip Lid Seal Dry Box Seal and Tank Assembly onto the new Bot tom Flip Lid Assembly as instructed in the previous sections 3 Place a new warning label and Seri...

Страница 134: ...AP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 6 Install heated tube wire connector into ISO Port and verify it is connected the whole way as shown below 7 Install the ISO Port into the top openi...

Страница 135: ...RVICE TECHNICAL REFERENCE MANUAL 8 Verify the ISO Port is installed properly and flush with the ISO Port Cover 9 Verify correct orientation of ISO Port assembly as shown below port facing toward front...

Страница 136: ...ATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL 10 Open the Flip Lid and ensure there are no pinched wires 11 Route the wires flat as shown and ensure they are in the middle within the gr...

Страница 137: ...s listed below Qty 1 ea Serial RS232 Cable DB9F DB9M 6 ft PN 1037268 Qty 1 ea USB Cable Type A to Micro B PN 1104765 Qty 1 ea Hose 22mm 18 in PN 1008198 Qty 1 ea 15mm Heated Tube PN HT15 Qty 1 ea Germ...

Страница 138: ...og and to reset the real time clock 9 3 Testing Environment Specifications The DreamStation devices must be tested within the following temperature specifications 41 F to 95 F 5 C to 35 C 9 4 Software...

Страница 139: ...IGNING ON do not include the quotations then hit Enter on the keyboard 4 Exit the Command prompt and reboot the PC 5 Locate and install the DreamStation Family Drivers Notes A Windows Security prompt...

Страница 140: ...9 5 Final Testing Procedure 1 The following icon will be created on your desktop Launch the Test Executive from your desktop by double clicking on the icon 2 A window will popup asking for a user name...

Страница 141: ...SERVICE TECHNICAL REFERENCE MANUAL 4 Select the DreamStation ASV_NIV Service Test seq file then select OK to load it into the test execu tive 5 You will see the test files loaded into the software 6 C...

Страница 142: ...8 Follow the prompts and enter in the device identification information serial model number etc 9 Follow the next prompt by connecting all hardware items listed and as shown below and then select OK T...

Страница 143: ...complete the following prompt will be displayed 12 Connect the Humidifier and Heated Tube as shown below then select OK 13 Continue to follow all prompts until the test is complete 14 If the test fai...

Страница 144: ...PAGE 144 1127179 VER 02 DREAMSTATION BIPAP S T AND AVAPS SERVICE TECHNICAL REFERENCE MANUAL...