amount of pressure relief, with higher numbers providing addi-
tional relief. The default setting of Off.
Note: If you do not lock the Bi-Flex setting, the patient has ac-
cess to the setting and can adjust it from 1-3. They cannot dis-
able Bi-Flex.
Note: Bi-Flex is available up to 25 cmH
2
O in S mode.
Flex Lock
This enables you to lock the Flex setting if you do not want the
patient to change it.
Rise Time
Rise time is the time it takes for the device to change from the
expiratory pressure setting to the inspiratory pressure setting.
This screen allows you to adjust the rise time so you can find
the desired setting. A setting of 1 is the fastest rise time, while
6 is the slowest.
Rise Time Lock
This enables you to lock the Rise Time setting. Select Off to
allow the user to adjust the Rise Time setting from 1-6. Select
On to lock the user from adjusting the setting.
Tube Type
This setting allows you to select the correct size diameter
tubing that you are using with the device. You can choose
either (22) for the Philips Respironics 22 mm tubing, or (15) for
the Philips Respironics 15 mm tubing. When using Heated
Tubing, the device will automatically change this setting to the
appropriate tubing type (15H).
Tube Type Lock
This enables you to lock the Tubing type setting for either the
15 mm or the 22 mm tubing if you do not want the patient to
change it.
Mask Type
This setting allows you to select the appropriate Mask Type
resistance setting (also known as System One Resistance
Control) for your Philips Respironics mask. This feature allows
the device to adjust the level of pressure compensation to
match your mask. Refer to the packaging of your mask to
identify the resistance setting for your mask.
Note: It is important to use the appropriate “Mask Type”
resistance setting to ensure proper pressure delivery to the
patient.
Mask Type Lock This enables you to lock the Mask Type resistance setting if
you do not want the patient to change it.
Check Mask Fit
You can enable or disable the check mask fit setting. This
feature allows the patient to check the fit of their mask prior to
starting therapy. This is done by measuring the amount of leak
in the patient circuit.