Code
Error Name
Error Level
Failed
Component
Actions
E-173
ERR_BT_STACK_MASK_
DATA_WR_FAIL
CONTINUE
PCA
1. Replace PCA and test.
E-174
ERR_BT_RADIO_EN_DE
N_TX
CONTINUE
PCA
1. Replace PCA and test.
E-175
ERR_BT_RADIO_EN_DE
N_RX
CONTINUE
PCA
1. Replace PCA and test.
E-176
ERR_BT_RADIO_BIST_F
AIL
CONTINUE
PCA
1. Replace PCA and test.
E-180
ERR_NVRAM_REMINDE
R_LOG
CONTINUE
N/A
1. Clear error log and test.
E-181
ERR_NVRAM_MUTEX_N
OT_AVAILABLE
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and test.
E-182
ERR_NVRAM_WRITE_FA
ILED
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and
test.
E-183
ERR_NVRAM_WRITE_S
U_FAILED
REBOOT
PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace PCA and
test.
E-190
ERR_LCD_ID_ERROR
REBOOT
1. UI Panel
2. PCA
1. If error was NOT Last
Stop Error, clear error log
and test.
2. If error was Last Stop
Error, replace UI Panel and
test.
3. If retest fails after UI
Panel replacement, replace
PCA and retest.