Code
Error Name
Error Level
Failed
Component
Actions
E-27
ERR_MOTOR_RL_NOCO
NVERGE
STOP
1. Motor
connection
2. Blower Box
3. PCA
1. Reseat motor connector,
clear error log and test.
2. If error still occurs,
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
E-28
ERR_NEGATIVE_QUADR
ATURE_VOLTAGE_VECT
OR
REBOOT
1. Motor
connection
2. Blower Box
3. PCA
1. Reseat motor connector,
clear error log and test.
2. If error still occurs,
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
E-29
ERR_VBUS_GAIN_ZERO
REBOOT
PCA
1. If E9 was not recorded,
clear error log and test.
2. If E9 was recorded,
replace PCA and test.
E-30
ERR_MOTOR_SPINUP_F
LUX_HIGH
REBOOT
1. Motor
connection
2. Blower Box
3. PCA
1. Reseat motor connector,
clear error log and test.
2. If error still occurs,
replace blower and retest.
3. If 2nd retest fails, replace
PCA.
E-34
ERR_MOTOR_TYPE_UN
KNOWN
STOP
PCA
1. Replace PCA and retest
E-35
ERR_MOTOR_BLOCKED
_INLET
CONTINUE
1. Air Path
2. PCA
3. Blower Box
1. Clear air path including
filter and test.
2. Load latest software
version if not up-to-date.
3. If test fails, replace PCA
and retest.
4. If retest fails, replace