Replacing the UI Panel
Kit
Included in Kit
Tools Required
UI Panel Assembly
•
UI Panel
•
Torx Screwdriver (T10 is
recommended)
To remove the UI Panel:
1. Remove all components as instructed in the previous sections.
2. Lift the UI Panel away from the device.
Note:
the alarm ribbon cable may detach when you pull the UI Panel away from the device. If it does
not detach, remove it by hand by gently lifting up on the ribbon.
3. Detach the main UI Panel ribbon from the PCA by lifting up on the connector latch and pulling the rib-
bon away from the connector.
To install the UI Panel:
1. Align the UI Panel ribbon into the connector latch on the PCA.
2. Snap down the connector latch to hold the ribbon in place.
3. Verify the UI connection is properly seated.
4. Attach the alarm ribbon cable to the PCA.
Note:
you may seat the UI Panel, and then attach the cable to the PCA if you find it easier.
5. Seat the UI Panel into Bottom Enclosure grooves (see illustrations below).
6. Assemble the remainder of the device as instructed in the previous sections.
Alarm Ribbon
Cable
Main Ribbon
Cable
(unclasp)