Alarm
Priority
Device Action
User Action
High
Operates
This alarm occurs when the
patient circuit is disconnected
or has a large leak. Press the
control dial to silence the
alarm. Press the control dial
again to acknowledge and re-
move the alarm from the
screen. Reconnect the pa-
tient circuit or fix the leak.
Device inop-
erative
Shuts down
Press the control dial to si-
lence the alarm. Disconnect
the device from power.
Device inop-
erative
Shuts down
Check device air inlet is not
obstructed. Check air filter(s)
are installed properly; replace
if needed.
Device inop-
erative
Shuts down
Check tube is not crushed or
folded such that air flow is re-
stricted. Check mask is at-
tached properly and without
any obstruction.
Device inop-
erative
Shuts down
Remove your mask, and re-
move power from the device.
Rinse or replace the reusable
air filter or replace the dispos-
able ultra-fine filter. Power
device on and resume normal
usage.